T- 2.6.5 NO BURNING AT R10 AND R11 / ISOLATE HEATER SWITCH
A)
Set your volt meter to RESISTANCE (Ω)!
B) Remove the Power Control board from its white clips.
C) Lay the power supply panel down to see the rear (solder) side of the board.
D)
PER the Figure below
, TWO (2) measurements:
Measurement #1:
BETWEEN ST13 AND ST7. Good = less than 2
.0 Ω. “OL” = way more than 2.0Ω!
Measurement #2:
BETWEEN ST5 AND ST4. Good = less than 2
.0 Ω. “OL” = way more than 2.0Ω!
E) BOTH measurements LESS THAN 2.0
Ω
?
Yes
BOTH LESS than 2.0
Ω! The Heater Switch is OKAY! See procedure number
281).
No One OR both MORE THAN 2.0
Ω! The Heater Switch is bad!