
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 415
Yes
Cables were OKAY! See procedure number
No
Problem with a cable! After the repair, perform parts a THROUGH d below:
a) Turn the machine OFF then allow even (7) minutes BEFORE continuing to part b.
b) Return to Dialysis Program (“Select Program”
→ ‘Dialysis’ → ‘CONFIRM’)!
c) Allow Conductivity to increase to more than 13.0 mS.
d) Call debug screen 1.Watch
FILACT
for two (2) minutes. If REMAINING = 0 the problem
is solved. If still going ever = 1 return to (ABOVE) procedure number
413).
FIL- 1.0.8 OLC ACTIVATED?
From debug screen 0, once again, does the OLC
POST Dialyzer Cond
symbol appear?
Yes
POST Dialyzer Cond
symbol appears! A problem with Post Dialyzer Cond. Close the shunt
door THEN proceed to
page 379
, p
No The OLC option will have to be activated as a bad OLC reading may cause Filling Programs EVEN
IF the OLC option is not activated! Perform parts a THROUGH i below:
a) Enter Service Mode
→ Options → Hardware Options.
b) Next to
OLC
(upper right) place the ‘X’ in the “Yes” box and press
‘CONFIRM’. The ‘X’ turns blue.
c) Turn the machine off then back on.
d) Return to Dialysis Program (“Select Program”
→ ‘Dialysis’ → ‘CONFIRM’)!
e) Call debug screen 0 and confirm the Post Dialyzer OLC symbols now appear.
f)
The shunt door MUST be open!
g) From the Home screen, ENSURING [Dialysate Flow] is set at 800 ml/min, allow
Conductivity
to reach between 13.0 and 14.5 mS!
h)
IMPORTANT!
Allow two (2) additional minutes for Post Conductivity to stabilize!
i)
Call debug screen 0. Are PRE AND POST Cond within 0.4 of each other now?
Yes
PRE AND POST Conductivity within 0.4 mS of each other! See procedure
number
(page 416).
No PRE AND POST Conductivity ARE NOT within 0.4 mS of each other! A problem with
Post Dialyzer Cond. Close the shunt door and proceed to
page 379
, p
rocedure
(page 379).
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...