
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 41
F- 3.1.1 FLOW LESS THAN 600 / ISOLATE ACTUATOR-TEST BOARD / VALVE #43
a) ENSURE
FILACT
REMAINS = 1!
b)
Figure below
, using a flashlight, check through Valve #43’s OUTPUT tubing for bio-growth
restrictions.
If a restriction is located this may be the problem!
Figure 8 – Hydraulics TOP View
c)
BEFORE continuing to part d allow four (4) minutes for Valve #43’s solenoid to
become warm!
d) Touch Valve #43’s (black) solenoid. Is it warm (Yes or No)?
Yes
Warm solenoid! TWO (2) possible bad components:
1)
Actuator-Test
*
board OR;
2
) Valve
#43.
*
To LOCATE the Actuator-Test board refer to
(page 9).
No
The solenoid in NOT warm!
NOTE
ONLY
VALVE #43
will be checked and proceed to
page 192
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...