
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 585
DIASAFE- 3.0.0 PRESSURE TESTS PASS
A) Figure right, ENSURE Valve #28 is plugged
PROPERLY into distribution board position,
“V28”!
B)
Figure below
, inside the hydraulics and the
rear side of the Diasafe
®
housing, ENSURE
‘TUBE A’ goes to VALVE #24 AND ‘TUBE B’
goes to VALVE #26!
C) Are you ABSOLUTELY SURE the DiaSafe
®
tubing is plumbed correctly?
Yes
Tubing is plumbed properly! Continue to part D.
No Tubing is backwards! This most likely is the problem! Per the Figure above turn the machine
off and attach the tubing properly.
D) From the Home screen, allow [
Conductivity
] to stabilize between 13.2 and 14.4 mS.
Parts F through I next page
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...