2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 305
T- 7.2.1 TEMPERATURE CONTROL
a) ENSURE the concentrate connectors are attached to acid and
LIQUID
bicarb!
b) From Service Mode
→ Calibrate Sensors → Temp Control.
c) ‘Sharply’ press ‘CONFIRM’. The screen should say “
3.
Connect an external temperature meter to the dialysate
lines”.
d) Install a Temperature (°C) meter into the dialysate lines.
Figure right, flow through the meter MUST be
bottom to top.
e)
IMPORTANT! Close the shunt door!
f) ‘Sharply’ press ‘CONFIRM’. The screen says
“5. Adjust the TEMP
DAC…”
and the
[TEMP DAC]
data box (Figure right) turns pale yellow /
white!
g) Based on the
[TEMP DAC]
data box, TWO (2) possible scenarios:
1)
IF (and ONLY if) between 120 and 180:
(page 306).
2) IF NOT between 120 and 180:
Perform parts a through d below:
a) Select
[TEMP DAC]
, it turns bright yellow.
b) Set
[TEMP DAC]
to “140”.
c) ‘Sharply’ press ‘CONFIRM’ ONCE.
[TEMP DAC]
MUST be pale yellow/white (NOT gray).
(page 306).
LEFT BLANK INTENTIONALLY
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...