2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 236
T- 1.0.28 METER VERSUS THE MACHINE’S [TEMPERATURE] DISPLAY IN RANGE
Proceed according to why you ORIGINALLY started troubleshooting. THREE (3) possible scenarios 1) or 2)
or 3) below:
1)
IF the problem was in a Heat Disinfect AND a “TEMP OVER 95 DEGREES” occurred:
Proceed to
page 246
, procedure number
2) IF the problem was in a Heat Disinfect AND [Temperature] did NOT reach 80° C:
Proceed to
page
, procedure number
3)
IF the problem was in Dialysis Program:
Per the Figure below, TWO (2) possible scenarios:
Scenario #1: IF NOT equipped with the
Blood Temperature
Module (BTM):
A
temperature problem IS
NOT occurring at this time.
Scenario #2: IF equipped with the
Blood Temperature
Module (BTM):
Allow the
machine to run for several
hours to see if a
temperature problem
reoccurs. TWO (2)
possible scenarios:
1)
IF the temperature problem reoccurs
The BTM is okay! See (ABOVE) procedure
page 221
2) IF a temperature problem DOES NOT reoccur:
See parts a THROUGH i below:
a)
Place the machine into Service Mode → Options → Hardware Options.
b) Next to ‘
BTM
’, place the ‘X’ into the “Yes” box then press ‘CONFIRM’. The “X”
turns blue!
c)
To prevent damage, turn the machine OFF!
d) Open the card cage.
e) For part f, the Functional board is the 3
rd
board from the right!
Parts f through i next page
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...