2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 168
Yes More than 200 ml! Valve #39 is okay! See procedure number
(page 168).
No Less than 200 ml! Valve #39 may be bad! See parts a AND b below:
a) ENSURE Valve #39 is plugged properly into distribution board position V39!
b) Swap in the following components (see Component List below) one at a time, with known
good, then repeat (ABOVE) procedure number CLEAN- 2.0.3 (page 167). Whhen more
than 200 ml is collected the last component swapped in is the problem.
COMPONENT LIST (4 components):
1)
Actuator-Test Board;
2)
Valve #39 (located on
the bottom of the hydrochamber, refer to the Figure above), including its blue wiring
harness;
3)
ACTUATOR board ribbon cable;
4)
Distribution board.
CLEAN- 2.0.4 MORE THAN 200 ML PER MINUTE/ VALVE #39 IS OKAY
If all procedures were performed correctly you have determined good flow through Restrictor #48 AND
Valve #39 i.e. more than 300 ml per minute to the Deaeration Pump. See parts a and b below:
a)
Remove the clamp from Restrictor #48’s tubing!
b) Return to
page 165
, procedure number