
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 513
BL- 2.0.0 BLOOD LEAK ALARM PRESENT (CHECK FOR AIR)
IMPORTANT!
Air in dialysate cause false blood leak alarms. Per the Figure below,
using a flashlight
,
watch for air bubbles through the Blood Leak Sensor’s TO AND FROM tubing for one (1) minute each?
Yes
Air seen! See procedure number
(page 514).
No air seen! See procedure number
page 516
).
Figure 89 – Hydraulics Side View
LEFT BLANK INTENTIONALLY
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...