2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 553
SECTION 21 – HYDRAULIC LEAKS
NOTE: T
hese procedures do NOT troubleshoot Vent Tubing overflow!
A)
IMPORTANT! Turn the machine OFF!
B) ENSURE the ‘Incoming Water’ and ‘To Drain’ tubing (Figure right)
are not reversed connected!
C) ENSURE the ‘To Drain’ tubing is not kinked AND, if using a drain
’Quick Connector’, it is attached PROPERLY to the station drain!
D) Reattach any ‘blown off’ tubing and dry the area!
E) Until
INSTRUCTED OTHERWISE
, direct the Fluid Sample
Connector (Figure right)
into a bucket!
F) Per the Figure below, was tubing ‘blowing off’ a UF Check
Valve?
Yes UF Check Valve tubing ‘blowing off’! See procedure
No UF Check Valves okay! See procedure number
page 555
HYDRAULICS TOP VIEW