2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 536
Yes
Between 35 and 36 psi! See procedure number
(page 536).
No
Is NOT between 35 and 36 psi! ENSURING the machine was in RINSE AND no leaks, TWO
(2) possible scenarios:
1)
IF (and ONLY if) pressure is too low:
DO NOT calibrate instead proceed to
page
.
2) IF pressure is too high!
Perform parts a AND b below:
a) Adjust Valve #78
*
until pressure cycles to between 35 and 36 psi!
*
To LOCATE #78
refer to
(page 535).
b) Leaving the machine in RINSE, see procedure number
(page 536).
BC LEAK- 2.0.0 DIAPHRAGM LEAK TEST
Two (2) buckets are required, one empty and the other filled with one gallon of
ACID CONCENTRATE
.
Place both buckets on the floor and read the procedure before continuing:
a) Call debug screen 5. Allow
FPRE
(middle column, ABOVE
FPOS
) to reach LESS THAN two thousand
eight hundred (2800). This verifies all conductive fluids have been rinsed out.
b) ENSURING RINSE is running (i.e. Remaining Time is NOT =
0:00) open the shunt door.
c) Figure right, direct the Ultrafiltrate Sample Connector tubing
into the EMPTY BUCKET to prevent UF circuit pressure.
d)
IMPORTANT!
Per the Figure
below, TIGHTLY clamp the
tubing at Valve #29. Ignore Flow Errors for now!
Parts e through g next page
Figure 98 – Valve #29 Location
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...