2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 136
MOTORS- 4.6.0 ISOLATE MOTOR VOLTAGE
a) Figure right, avoiding the VACANT positions on the left
and right, return the motor connector,
without the
cap
, to its distribution board position.
b) ENSURE the connector is aligned properly!
c) Set your CALIBRATED volt meter to
DC voltage
(V
DC
)
.
d) Measure,
inside the distribution board
, between the
terminals where the wires are soldered.
IMPORTANT!
ENSURE good contact with BOTH terminals!
below, perform the indicated Response based on measured voltage.
Table 4 – Motor Troubleshooting
MOTOR
DISTRIBUTION
BOARD
POSITION
DC VOLTAGE
READING
Your RESPONSE
DEAERATION
P20, “DEGAS-P”
15 volts DC or more
TWO (2)
possible bad components
:
1)
Bad deaeration motor OR
2)
Intermittent 24V Power Harness
connection. See Figure below.
Less than 12 volts DC
Bad motor voltage! See procedure number
(page 137)
FLOW
P21, “FLOW-P”
5 volts DC or more
TWO (2)
possible bad components
:
1)
Bad flow motor OR
2)
Intermittent 24V Power Harness
connection. See Figure below.
Less than 5 volts DC
Bad motor voltage! See procedure number
(page 137)
24V Power Harness (Behind the card cage / Inside the Power Supply)
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...