2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 260
T- 1.8.0 ISOLATE NTC #2
a)
Turn the HEATER Switch OFF!
b) Figure right, move NTC #2’s connector from the
1
st
distribution board position from the left, “CON-NTC” to
the
2
nd
position from the left, “MON-NTC”, AVOIDING
the VACANT position at the right,
c) Call debug screen 4. Allow up to seven (7)
minutes
OR
until if
TEMP
(lower left) falls to
2.5 OR less?
Yes
TEMP
= 2.5 OR less! NTC #2 appears to be okay. See procedure number
No
TEMP
more than 2.5! See parts a and b below:
a) ENSURING the HEATER Switch was OFF, return NTC #3’s connector to
2
nd
distribution
board position from the left, “MON-NTC”!
b) NTC #2 is bad! To LOCATE NTC #2 refer to
(page 218).
T- 1.8.1 NTC #2 OKAY / VERIFY ‘TEMP COMP’ OPTION
a)
Return NTC #2’s connector to 1
st
distribution board position from the left, “CON-
NTC”!
b)
Avoiding the VACANT position at the right (“PH-PR”), return NTC #3’s connector
to the 2
nd
distribution board position from the left, “MON-NTC”!
c)
Turn the HEATER Switch ON!
d)
Enter Service Mode → Options → Hardware Options. Does the “Yes” box next
to
‘Temp Comp’
have a blue ‘X’ in it (Yes or No)?
Yes
Place the ‘X’ in the “No” box and press ‘CONFIRM’. The ‘X’ turns
blue. Post a note that ‘
Temp Comp
’ is off then see procedure
(page 261).
No Temp Comp = No! See procedure number
(page 261).
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...