2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 425
FIL- 3.0.0 FILACT = 0 / CREATE VENOUS PRESSURE
This procedure creates Venous Pressure to ensure Valve #25 remains open for the subsequent test:
a)
Turn Dialysate Flow OFF (Dialysate Flow on/off lamp blinks)!
b) Call debug screen 0. ENSURE all eight balancing chamber valves (#31 through #38) ‘dots’ are
REMAINING white (Flow is Off)!
c)
IMPORTANT! Dialysate Flow MUST REMAIN OFF till instructed otherwise!
d)
Figure below
, attach a syringe, with a piece of tubing attached that will fit SNUG to the Level
Detector’s INNER P
ven
port.
e) Call debug screen 1.
f) Push on the syringe plunger until
VEN
(middle column) is between 400 and 450.
g) Clamp the syringe tubing to keep the pressure!
h) If
VEN
falls more than three (3) mmHg in thirty (30) seconds there is a leak at the P
ven
port OR inside
the module.
i)
Leaving the clamp in place, to keep
VEN
more than 400, see procedure number
(page 426).