2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 485
PHT- 4.2.0 LOADING PRESSURE OKAY
a) Turn the water off and allow a “No Water” alarm to
occur.
b) The Deaeration gauge is used next. ENSURE it reads
0 inHg BEFORE installing it!
c) Allow one (1) minute for the deaeration motor to stop.
d) Figure right, tee the gauge to the Inlet (clear tubing)
of the Deaeration Pump.
e)
IMPORTANT! Turn the water on and allow
the “No Water” alarm to go away!
f) Deaeration Pressure OKAY? Refer to
(page 667) for what pressure should be.
Yes
Deaeration pressure is OKAY. See procedure number
(page 485).
No
Deaeration Pressure in NOT OKAY! ENSURING a “No Water” alarm is NOT presenting,
NOTE this page number, as you will return here, THEN proceed to
page 498
PHT- 4.3.0 DEAERATION PRESSURE OKAY
a) Turn the water off and allow a “No Water” alarm to occur.
b) Allow one (1) minute for the deaeration motor to stop.
c) Remove the gauge and reattach the tubing!
d)
IMPORTANT!
Turn the water on and ENSURE the “No Water” alarm goes away!
e) Allow the machine to come out of bypass i.e. the external flow indicator’s ‘bob’ moving up and down in
the sight glass!
f) Allow one (1) more minute BEFORE continuing to part g.
g) Reset all alarms then select the
[Test & Options]
tab. TWO (2) possible scenarios.
1)
IF the [Pressure Test] button is BLUE:
Select the
[Pressure Test]
button THEN press
‘CONFIRM’ THEN IMMEDIATELY see procedure number
(page 486).
2) IF the [Pressure Test] button is GRAY:
Select
[Both Tests]
THEN press ‘CONFIRM’ THEN
IMMEDIATELY see procedure number
(page 486).
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...