2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 456
TMP- 2.0.16 ISOLATE TMP
a)
Turn Dialysate Flow OFF i.e. Flow on/off lamp blinks!
b) Call debug screen 0. If (and ONLY if) ALL EIGHT (8) Balancing Chamber Valve ‘dots’ (#31 - #38) are
REMAINING WHITE i.e. Dialysate Flow is OFF see part d! NOTE: If the Balancing Chamber Valves
are NOT remaining white allow debug screen 1’s
FILACT
= 0 then they will!
c)
Leaving Flow OFF till instructed
(page 456).
TMP- 2.0.17 ISOLATE VENOUS PRESSURE
This procedure creates Venous Pressure (
VEN
) to keep Valve #25 open for subsequent tests!
a) Figures below, attach a syringe with a piece of tubing attached that will fit SNUG to the Level
Detector’s INNER P
ven
port.
b) Call debug screen 1. Push on the syringe plunger until
VEN
(middle column) is
between 400 and 450.
c) Clamp the syringe tubing to HOLD the pressure!
d) If
VEN
falls more than 3 mmHg in thirty (30) seconds there is a leak at the P
ven
port OR inside the
module.
e) Leaving the clamp in place to keep
VEN
more than 450 see procedure number
(page 457).