2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 209
VEC- 1.3.0 ISOLATE DRAIN VALVE #30
a) Turn the machine OFF then back on. Allow the “Select Program” screen to appear but
DO NOT
press any keys till instructed!
b) With “Select Program” up, Figure right, screw a 60 ml
syringe, filled COMPLETELY with water, onto the
Fluid Sample Port.
c) Start a SEVEN (7) second timer, in your head, as you
begin to push
as HARD as you can
on the
syringe plunger! You should feel very little resistance!
d) Can you push ALL of the water through the Sample
Port within SEVEN (7) seconds?
Yes
Good flow through Valve #30! See procedure
number VEC- 1.4.0 (page 209).
No
Significant resistance!
NOTE
ONLY
VALVE
#30
will be checked and proceed to
page
VEC- 1.4.0 GOOD FLOW THROUGH VALVE #30 / ISOLATE VALVE #24
a) Obtain a 1000 ml graduated cylinder.
b)
Return to Heat Disinfect OR Rinse!
c) Open the shunt door and
LEAVE IT OPEN TILL INSTRUCTED!
d) Remove the BLUE dialyzer quick connector from the shunt but
DO NOT CLOSE THE
DOOR!
The Program MUST NOT Interrupt!
e) Measure from the blue dialyzer connector, for thirty (30) seconds! More than 300 ml collected?
Yes
More than 300 ml! Without returning the connector to the shunt yet, see procedure number
(page 211).
No
Less than 300 ml! ENSURING the machine was in Heat Disinfect OR Rinse see parts a
THROUGH c below:
a) Return the dialyzer connector to the shunt but
DO NOT
close the door.
Parts b and c next page
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...