2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 211
VEC- 1.5.0 VALVE #24 OKAY / ISOLATE VALVE #25
a) With the shunt door open, the dialyzer connector NOT in the shunt AND still in Heat Disinfect OR Rinse,
ENSURE the external flow indicator’s ‘bob’ is rising!
b) Return the dialyzer connector to the shunt but
DO NOT
close the door!
c) Watching for thirty (30) seconds, does the external flow indicator’s ‘bob’ rise at least ½ way up now?
Yes
Bob moving! Something changed!
A)
Turn the machine OFF;
B)
Close the door;
C)
Turn
the machine on and return to Heat Disinfect OR Rinse;
D)
If a Flow Error reoccurs return to
(page 159).
No ‘Bob’ not moving!
NOTE
ONLY
VALVE #25
will be checked and proceed to
page 192