2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 159
CLEAN- 1.2.1 PREPARE TO ISOLATE LOADING PRESSURE
a)
Figure right
, remove the ACID (RED)
connector from the acid inlet tubing AND
attach the Loading Pressure gauge tubing to
it.
b)
ENSURE
the gauge reads 0 psi before
inserting it in the Rinse Port!
c) Place the Acid Inlet tubing into water.
d) Figure below,
SLAM
the acid connector,
with the gauge now attached, into the
Acetate/Acid Rinse port.
(page 159).
CLEAN- 1.2.1.1 ISOLATE LOADING PRESSURE (1)
a) If the “Interrupted” banner (Figure right) IS NOT up
continue to part b. If “Interrupted” is up press
‘Escape’, then depending on where the Flow Error was occurring, return HEAT DISINFECT OR RINSE
!
b) ENSURE no leaks at the gauge!
(page 160).