2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 131
MOTORS- 3.0.0 MOTOR ROTATING COUNTERCLOCKWISE (CCW)
a) To ENSURE a proper connection, per the Figure below, trace the wires from the
NOTED
motor to the
Distribution B
oard. Deaeration Pump → Distribution Board position “P20”; Flow Pump → distribution
board position “P21”.
b) To ENSURE the pump head is oriented correctly, per the
below, the ‘ID Decal’, MUST be
either on the TOP OR FRONT of the pump head. If on the front, it MUST be right side up!
c) If not already, place the machine in DIALYSIS PROGRAM!
d) Open the shunt door and remove both dialyzer connectors!
e) CLOSE THE DOOR then call debug screen 2 to see
! EMPTY
(left column). TWO (2)
possible scenarios:
1)
IF (and ONLY if) ! EMPTY = 1:
2)
IF ! EMPTY = 0:
Either the (red) arterial dialysate connector is NOT connected to the shunt door
properly OR FOUR (4) possible bad components:
1)
Bad Actuator-Test Board OR;
2)
Bad
Functional Board OR;
3)
Bad arterial and/or venous line shunt door switch OR;
4)
Bad motherboard.
*
Swap in each component one at a time, with known good, until
! Empty
= 1
** To prevent “Cond Offset Failure”, place the machine into
T and C Mode
(refer to
OPERATING
MODES
, page
Error! Bookmark not defined.
Figure 29 – Deaeration and Flow Motor Location / Orientation
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...