2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 403
e) Allow thirty (30) seconds THEN, WITHOUT LOOKING AWAY, watch the pump’s symbol for two (2)
minutes. Does it remain white?
Yes
Pump symbol remains white! The previous pump is bad. Repeat procedure number
If BOTH symbols remain white:
BOTH previous
pumps are bad.
If one pump is pink and the other is white:
See (ABOVE) procedure
(page 397).
No Pump symbol pink! FIVE (5) possible bad components (referring to
1)
Bad actuator cable OR;
2)
Bad Sensor Board cable OR;
3)
Bad Sensor Board OR;
4)
Bad
distribution board OR;
5)
Bad mother board