2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 639
P- 2.0.1 SYSTEM INITIALIZATION DOES NOT REACH 100%
a)
IMPORTANT!
Turn the machine OFF!
b)
IMPORTANT!
Reconnect the ACTUATOR cable.
c)
Figure below
, behind the card cage, unplug the
ALL MODULES ONLY
including the Blood
Pressure module!
d)
From here forward,
ignore “Art BP No Comm” OR “Blood Still Sensed!” if they occur!
e) Turn the machine ON (fan running). Does System Initialization
reach 100%?
Yes
System Initialization reaches 100%! See
ISOLATING THE MODULES
this page
No
System Initialization does NOT reach 100%! ENSURING the ACTUATOR cable has been
reconnected, see procedure number
(page 640).
ISOLATING THE MODULES
A)
Turn the machine OFF!
B) Your choice, CAREFULLY* plug one of the modules back in. *
NEVER reverse connect the modules!
NEVER plug the Blood Pressure module in with the machine on!
C) Turn the machine on (fan running).
D) Repeat parts A through D until System Initialization does NOT reach 100% indicating the last module
plugged in (or its cable) is causing the problem. If after plugging in all modules and System Initialization
reaches 100% the problem may be heat related and requires running the machine in Heat disinfect to
locate it.
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...