2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 176
CLEAN- 4.3.0 VALVE #29’S ‘DOT’ ALWAYS BLUE
Call debug screen 1 (Figure right). Does
VERR
(lower right column) = 0?
Yes
VERR
= 0! Proceed to
page
, procedure number
No
VERR
= 1 OR more! See procedure
(page 176).
CLEAN- 4.3.1 VERR = 1 OR MORE
a) Call debug screen 0
(Figure right).
b) Ignoring the top F
low
Error
window,
in part c,
you will watch the
2
nd
window down,
Valve
Error
.
c) Watching for one (1) minute, ignoring a ‘blink to 1’ that lasts less than one (1) second, does
Valve
Error
EVER = 1 for LONGER THAN two (2) seconds?
Yes
Valve Error
EVER = 1 LONGER THAN two (2) seconds! Proceed to
page 189
No
Valve Error
= 0 OR ‘blinks to 1’ for less than one (1) second!
Close the shunt door THEN
proceed to
page 206
TROUBLESHOOTING VALVE ERRORS IN CLEANING
.
LEFT BLANK INTENTIONALLY