2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 331
CO- 1.0.8.1 [CONDUCTIVITY] MORE THAN 14.5 mS / ISOLATE PUMP TUBING CONNECTIONS
a) Per the top Figure below, trace the tubing from the red (ACID) connector to ENSURE it goes to the
input of the
ACID PUMP
b) If ABSOLUTELY SURE the pump is plumbed correctly, see procedure number
(page 332).