2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 84
F- 7.6.10 PART 3 BAD FLOW / PART 4 MORE THAN 800 / RESTRICTED VALVES #36 AND/OR #32
Balancing Chamber Valve #36 and/or Valve #32 is not opening. Also, the Actuator Cable or the distribution
board may be bad. These procedures isolate between these components:
A) Turn the machine OFF!
B) Return valves #31 AND #38 to their PROPER distribution board positions.
C) The Flow Pump remains plugged into the Deaeration Pump’s distribution board position!
D)
Per the Figure below
,
replace
Valve #36
with a known good!
E) Plug the new valve #36’s connector into valve #33’s distribution board position, “V33”.
F) ENSURE the bucket with the RED dialyzer connector submersed is full!
G)
Turn the machine but DO NOT press any other keys!
H) With “Select Program” up, measure drain flow for one (1) minute. More than 800 ml every minute?
Yes
More than 800 ml! Problem located! The previous valve #36 is bad!
No Less than 800 ml collected! See procedure number
(page 84).
Figure 20 – Balancing Chamber Valves
F- 7.6.12 LESS THAN 800 ML COLLECTED / ISOLATE VALVE #32
A) Turn the machine OFF!
B) Return Valves #33 AND the new Valve #36 connectors to their PROPER distribution board positions!
C)
Per the Figure above
,
replace
Valve #32
with a known good!
Parts D through G next page
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...