2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 348
CO- 4.4.0 ISOLATE CONCENTRATE PICKUP SYSTEMS
a) Place the concentrate connectors
HARD
into their rinse ports!
b) Return to Rinse Program! ALLOW one (1) minute to prime the pumps.
c) Per the Figure (right), place the connector’s ‘adapter plug’
ONLY
into a
container of water i.e
. do
NOT submerse the whole connector!
d) Return to Dialysis Program
(“Select Program” → ‘Dialysis’ → ‘CONFIRM’)!
e) Is AIR FREE concentrate drawn steadily in one direction now?
Yes
Concentrate drawn! There is a problem with the pickup wand
(page 348).
CO- 4.4.1 ISOLATE THE CONNECTOR
a) Place the concentrate connectors
HARD
into their rinse ports
b) Return to Rinse Program!
c) ALLOW one (1) minute to prime the pumps
d)
Return to Dialysis Program (“Select Program” → ‘Dialysis’ → ‘CONFIRM’).
e) Per the Figure right, remove the connector from the inlet tubing and
submerse the tubing,
without the filter,
into a container of water.
f) Is AIR FREE concentrate being drawn steadily in one direction now?
Yes
Concentrate drawn! Per the Figure below, there is a problem with the concentrate connector.
(page 349).
Figure 59 – Concentrate Connectors