2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 383
CO- 10.0.30 PROBLEM WITH POST DIALYZER COND
ENSURING the shunt door is open, is screen 0’s
Post Dialyzer Cond
more than 10.000 mS?
Yes
Post Cond more than 10.000! See procedure number
No
Post Cond less than 10.000! See procedure number
CO- 10.0.32 POST DIALYZER COND MORE THAN 10.000
Is screen 0’s
Post Dialyzer Cond
more than 20.000 mS?
Yes
Post Cond more than 20.000! See procedure number
No
Post Cond less than 20.000! Proceed to
page 354
, procedure number
CO- 10.0.34 ISOLATE POST COND CIRCUIT
a) Figure right,
from the distribution board, unplug the
10
th
connector cap from the LEFT. This is Post
Dialyzer Cond Cell #13, “x13, COND-POS”!
b) Using a flashlight, check inside the “x13” position.
If
damage or corrosion is located this may be the
problem!
c) Call debug screen 5, Is
FPOS
more
than 4000 (four thousand) [Yes or
No]?
Yes
FPOS
more than 4000! Either Post Dialyzer Cond Cell #13 was not unplugged from
position “x13” or there is a problem with its circuit. THREE (3) possible bad
components:
1)
Sensor Board
1
;
2)
Sensor Board cable;
3)
distribution board.
1
To prevent “Cond Offset Failure” place the machine into T and C mode. Refer to
OPERATING MODES
, (page
Error! Bookmark not defined.
).
No
FPOS
less than 4000! See parts a AND b below:
a) Per the Figure right, place the
274 Ω
resistor plug, from the
TWO-RESISTOR SET
, into the Post Dialyzer Cond Cell’s #13
vacant distribution board position, “x13, COND-POS”.
b) Is
FPOS
between 13500 and 14500?
Yes
FPOS
between 13500 and 14500! See procedure number
No
Either the
274 Ω
resistor plug was not placed properly or there is a problem
with the Post Dialyzer Cond Circuit! THREE (3) possible bad components:
1)
Sensor Board
a
;
2)
Sensor Board cable;
3)
Distribution board.
a
To prevent “Cond Offset Failure” place the machine into T and C mode.
Refer to
OPERATING MODES
(page
Error! Bookmark not defined.
).
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...