
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 597
P- B.0.0 ISOLATE POWER SYSTEMS
a) Figure right, ENSURE the
3
rd
distribution
board position from the left, “x4, PH-P”,
IS
VACANT! If NOT, this may be the problem!
b) To avoid pulling cables loose, GENTLY open
the card cage.
A below, at the rear of the card cage, ENSURE the 24V POWER harness is plugged in
securely with the orange wire to the left.
If not, this may be the problem!
B below, check the power plug, cord and strain relief for signs of burning or other
damage.
Replace a damaged plug, cord or strain relief!
e) With the machine on i.e. message displayed on the screen ENSURE the fan is running i.e. air from the
power supply rear vents. If not air from the vents the fans is bad and this allows excessive heat which
causes electronic failures!
f) Proceed to
page 599
, procedure number
Figure 113A – 24V POWER Harness
Figure 113B – Power Supply Rear View
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...