2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 613
P- G.5.4 SYMPTOM REOCCURS / ISOLATE HYDRAULIC SENSORS
a)
IMPORTANT! To prevent damage turn the machine OFF!
b) To isolate ALL hydraulic sensors at the same time, Figure below, unplug the Sensor board’s ribbon
cable!
c)
IMPORTANT! Turn the machine on (fan running)!
d) If possible go to
Dialysis Program
or RINSE (DO NOT go to Heat Disinfect)!
e)
IGNORE acid and / or bicarb pump EOS alarms or possible grinding!
f)
ENSURE the black lead remains connected to chassis ground!
g) Re-m12 and / or +5 volts (within range now?). If the symptom was intermittent call debug screen
1 to watch
5V Est
and / or
12V Est.
Allowing sufficient time, does the symptom you were troubleshooting
reoccur?
Yes
Symptom reoccurs! Proceed to
page 615
, procedure number
No Screen on and / or good voltage(s)! See procedure number
(page 614).