2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 37
F- 2.0.10 VALVE #24’S ‘DOT’ IS BLUE
a) Call the Home screen to ENSURE [Dialysate Flow] has REMAINED set to 800 ml/min!!
b) Does a leak develop? Refer to
Table 1
Does a leak develop?
Your Response:
Yes (a leak
develops)
See parts a THROUGH c below:
a) Turn the machine OFF!
b) If tubing has blown off tie wrap these segments to where they
belong!
c) Proceed to
page 112
, procedure number
No leaks
TWO (2) possible scenarios based on if the external flow
indicator’s ‘bob’ is rising, at least ¼ way up, in the sight tube.
1)
IF (and ONLY if) rising at least
¼
way:
Proceed to
page
, procedure number
2)
IF NOT rising at least
¼
way:
See procedure number
(page 37).
F- 2.0.11 VALVE #24’S ‘DOT IS BLUE BUT ‘BOB’ IS NOT MOVING
Call debug screen 1 to see
VERR
(right column, bottom). TWO (2) possible scenarios:
1)
IF (and ONLY if) VERR = 1 or more:
Proceed to
page 195
2)
IF VERR = 0:
Proceed to
page 112
, procedure number
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...