2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 483
PHT- 3.1.5 ALL OTHER SCENARIOS / PRESSURE TEST FAILED
Has the DiaSafe
®
filter been replaced in THIS Troubleshooting session?
Yes
The DiaSafe
®
filter has been replaced! Proceed to
page 440
No a) Replace the DiaSafe
®
filter preferably with one from another machine that is passing
Pressure Tests.
b) Place the machine into RINSE for five (5) FULL minutes!
c) Return to Dialysis Program (
“Select Program” → ‘Dialysis’ → ‘CONFIRM’).
d) From the Home screen, set [Dialysate Flow] to 800 ml/min and press ‘CONFIRM’!
e) Allow the machine to come out of bypass i.e. the external flow indicator’s ‘bob’ moving up
and down in the sight glass!
f) Allow one (1) minute BEFORE continuing to part g.
g) Reset all alarms and repeat the Pressure Tests. Do they BOTH pass?
Yes
Both Pass! The DiaSafe
®
filter MAY have been causing the problem
HOWEVER, IMPORTANT!
484).
No Pressure test(s) Fail! Proceed to
page 440