2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 457
TMP- 2.2.5 ACHIEVE NEGATIVE PRESSURE
a) Figure right, connect the Four-Way Assembly
(P/N 150034) to the dialyzer connectors.
b) Place the Four-Way at dialyzer level!
c) ENSURE a transducer protector IS NOT in
the ‘to syringe’ tubing segment!
d) If using a
NEO-2
attach to the + (top, red) port. If
using a
90XL
attach to the module’s Gauge Port!
e) DO NOT allow tension in the Four-Way tubing
segments!
f)
IMPORTANT! CLOSE THE SHUNT
DOOR!
g) Call debug screen 2. To ENSURE the shunt door is indicated
closed
CVRCLS
= 1!
h) PULL on the syringe plunger. Can you achieve negative (-) 250 +/- 5 mmHg on the meter?
Yes
-250 achieved! Clamp the ‘to syringe’ tubing segment to keep the pressure then see
procedure number
(page 457).
No CANNOT achieve -250 on the meter! ENSURE the transducer protector, at the meter, is not
wet OR consider replacing it! If OKAY, see procedure number
TMP- 2.2.6 NEGATIVE (–)250 ACHIEVED / NEGATIVE PRESSURE HOLDING TEST
Does -250 HOLD, +/- 15 mmHg, for one (1) minute?
Yes
-250 HOLDS! Proceed to
page 459
, procedure number
No -250 does NOT hold! See procedure number
(page 457).
TMP- 2.2.7X -250 COULD NOT BE ACHIEVED OR IT DID NOT HOLD / ISOLATE FOUR-WAY
a)
Figure right
, clamp BOTH Four-Way
dialyzer Line tubing segments.
b) Can you achieve negative (-)250 mmHg AND
does it HOLD (+/- 15 mmHg) for one (1) minute
now?
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...