2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 381
Yes
[Post-Temperature Reference]
more than 20! Without pressing ‘CONFIRM’ yet, see
(page 381).
No
[Post-Temperature Reference]
is less than 20! See parts a and b below:
a) ENSURE the
34° C
(6.808 K
Ω) plug is placed properly at position “x44, NTC-POST”. If
NOT, repeat (ABOVE) procedure number
(page 380) AND/ OR consider
using the
34° C
plug from another
FOUR-RESISTOR SET!
b) If SURE the plug is placed correctly proceed to
page 394
, procedure number
CO- 10.0.24 POST TEMPERATURE REFERENCE MORE THAN 20
‘Sharply’ press ‘CONFIRM’ ONCE. Does an “Error” banner appear?
Yes
“Error” banner appears!
Read before performing!
Turn the machine off then
CAREFULLY repeat (ABOVE) procedure number
(page 380). CONSIDER
using the
34° C
plug from another
FOUR-RESISTOR SET
. If the “Error” banner
reoccurs proceed to
page 394
, procedure number
No “Error banner does NOT appear. See parts a through g below:
a) The screen says
“2. Connect a 5.117 K ohm resistor…”
.
b)
Place the 41° C (5.117
KΩ) plug, from the
FOUR-RESISTOR SET
, into distribution board
position, “x44, NTC-POST”.
c) The screen’s [
Post-Temperature Reference]
box
will
be more than 150 if the 41° C plug is
placed properly.
d) Press ‘CONFIRM’. The screen says “
3. Calibration saved”
and the screen’s [
Post-
Temperature]
data box should = 41° +/- 0.1 °C.
e) Press ‘CONFIRM’ again to save the calibration.
f) Return to Dialysis Program
(“Select Program” → ‘Dialysis’ → ‘CONFIRM’)!
LEFT BLANK INTENTIONALLY
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...