2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 625
P- H.4.5 +24 VOLT- A TROUBLESHOOTING
a) Spread the card cage panels open then gently drop the front panel down.
b)
ENSURE the volt meter’s black lead remains attached to chassis ground!
c) ENSURE the machine in on (fan running)!
d)
below
, place the red meter lead into the RIGHT SIDE socket of the motherboard’s white, two
(2) pin, +24V-A connector. BE CERTAIN to make good contact with the metal pin inside the socket!
e) TWO (2) possible scenarios:
1)
IF between 23.0 and 28.0 volts DC (V
DC
):
(page 626).
2)
IF NOT between 23.0 and 28.0 volts DC (V
DC
):
Proceed to
page 649
, procedure number
Figure 119 – Mothe24V-A Connector