2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 649
P- 4.0.0 ISOLATE THE POWER SUPPLY
a)
CAUTION!
To prevent an electrocution
hazard unplug the machine!
b) Figure right, trace the cables between the
Acid, Bicarb AND UF Pumps and the
distribution board to ENSURE they are NOT
pinched or damaged!
c) Figure below, trace the cables between the
modules and the rear of the card cage to
ENSURE they are not pinched or damaged!
Parts d through j next page