2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 453
Yes
The UF ‘dot’ blinks blue! If SURE the green UF lamp was OFF, the Actuator-Test Board may
be bad!
No The ‘dot’ REMAINS white!
, remove the UF Pump’s output (TOP)
tubing! Look at the UF Pump’s nozzle! TWO (2) possible scenarios:
1)
IF (and ONLY if) NO fluid output from the nozzle! A)
Reattach the tubing!
B)
See
(page 453).
2)
IF fluid output from the nozzle!
TWO (2) possibilities:
1)
The UF Pump is upside
down! OR;
2)
Incorrectly installed OR bad UF Pump seals/springs.
Figure 70 – UF Pump OUTPUT
TMP- 2.0.7 NO UF PUMP LEAKS
a)
IMPORTANT! Remove the ‘dummy chamber’ from the Level Detector module!
b)
IMPORTANT! CLOSE the shunt door!
c) From debug screen 0, Figure right, allow Valve #24’s ‘dot’ to
REMAIN BLUE i.e. normal AND stable Temp, Cond AND shunt
door closed!
d) The external flow indicator’s ‘bob’ MUST be moving up and
down!
e) Call debug screen 1. Watch
FILACT
for thirty (30) seconds to
ENSURE it = 0 ALWAYS.
(page 454).