2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 95
F- 9.0.26 BAD FLOW PUMP PRESSURE
a) Return to Calibrate Hydr
aulics → Flow Pressure.
b) Press ‘CONFIRM’ THEN ENSURE the FLOW MOTOR shaft is rotating!
c)
Per the Figure below
, above the Deair Pump, locate Valve #78.
d)
WARNING!
To avoid error DO NOT use a plastic clamp in part e! Use METAL NEEDLE NOSE
PLIERS instead!
e) While watching the gauge,
using PLIERS
, clamp then QUICKLY unclamp the white tubing at the top
of Valve #78 several times.
f) When clamped, does gauge pressure reach forty (40) psi or more?
Yes
40 psi or more! The Flow Pump is okay! See procedure number
(page 96).
No Less than 40 psi! TWO (2) possible bad components:
1)
Flow Pump head OR;
2)
Flow Pump
motor (possibly brushes).