2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 137
MOTORS- 4.7.0 BAD MOTOR VOLTAGE MEASUREMENT / ISOLATE A ‘BINDING’ MOTOR
a)
IMPORTANT! Turn the machine OFF!
b) Figure right, remove the pump head from the
NOTED
motor to
expose the ‘Drive Magnet’.
c) Manually spin the ‘Drive Magnet’. Does it rotate freely i.e. is NOT
‘binding’?
Yes
Rotates freely!
See procedure number
(pa
No DOES NOT rotate freely! The motor is bad and brushes won’t help.
CAUTION!
The ‘binding’
motor may have damaged the Actuator-Test and/or Power Logic Board!
MOTORS- 4.8.0 ISOLATE BAD MOTOR VOLTAGE
a) Return the pump head to the motor!
b) Figure right, ENSURE the motor is plugged properly into
its CORRECT distribution board position!
c) Swap in the listed components (see
COMPONENT LIST
below), one at a time, with known good and in between
continue to parts d and e to test the new component!
COMPONENT LIST:
1)
Actuator-Test Board*; 2
)
Actuator board ribbon cable;
3)
Distribution board;
4)
Power supply connections.
* To LOCATE the board refer to
(page 9)
d) Return to either RINSE OR Dialysis.
e) Check for motor for rotation to see if the new component causes it to run. If not turn the machine off and
repeat parts c through e until it does run!
LEFT BLANK INTENTIONALLY
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...