2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 127
F- 18.1.0 ISOLATE LOADING PRESSURE VALVE #65
a) Ignoring the MINIMUM pressure for now, if Loading Pressure is NOT achieving a PEAK of somewhere
between 23 and 25 psi, adjust Loading Pressure Valve #65’s nut (Figure below) until it does.
b) Figure right, if threads are visible under Valve #65’s nut see
part c. If no threads are visible either the wrong spring is
installed under Valve #65’s nut* OR Valve #65 is bad.
* Refer to
(page 19)
c) From the Home screen, set [Dialysate Flow] to 800 ml/min!
d) TWO (2) possible TMP window scenarios:
1)
IF (and ONLY if) the TMP window is white:
See
(page 127).
2)
IF the TMP window is RED (TMP alarm present):
See parts a THROUGH c below
:
a) Press and release the ‘Reset’ key then immediately press and hold it for three (3) seconds.
b) Allow thirty (30) seconds. If a TMP alarm reoccurs attempt RESET up to twice more BEFORE
continuing to part c.
c) Allow thirty (30) seconds BEFORE continuing to procedure number
(page 127).
F- 18.2.0 ISOLATE POSSIBLE LEAKING BALANCING CHAMBER VALVE
Call debug screen 4 to simultaneously watch
PDIA
(left column) AND
ADIA
(right column) or two (2) FULL minutes. They may change slightly but
should remain between 2.0 and 8.0. TWO (2) possible scenarios next
page:
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...