2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 581
HEATER RELAY- 2.0.1 DIODE 17 IS GOOD
a) Reconnect the Power Control board’s cables.
b) Slide the power supply back into the cabinet but DO NOT mount it yet!
c) Turn the machine off and swap in a known good
Power Logic Board. To LOCATE the Power Logic board
refer to
(page 9).
d) Turn the machine on and return to Dialysis Program (
“Select Program” → ‘Dialysis’ → ‘CONFIRM’).
e) TWO (2) possible scenarios:
1)
IF “Heater Relay Test Fail” reoccurs:
Turn the machine off then see procedure number
(page 582).
2)
IF “Heater Relay Test Fail” does NOT reoccur:
Problem solved. The previous Power Logic Board
was bad.
LEFT BLANK INTENTIONALLY