2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 15
Yes
“Operator Error” occurs! Proceed to
page 618,
No
“Operator Error” DID NOT occur! See part N.
N) If done correctly, the screen says
“4. Verify that 5V EST is between 4.8 to 5.2….”.
O) Based on the
screen’s
[5V EST] AND [12V EST] windows, TWO (2)
CHECKS:
Check #1:
Is
[5V EST]
between 4.8 and 5.2?
Check #2:
Is
[12V EST]
between 11.7 and 12.3?
Yes (to BOTH):
Press CONFIRM to save the calibration THEN see part P.
No (to one OR both):
Proceed to
page 618
, procedure number
END OF ‘VOLTAGE DETECTOR CALIBRATION’