2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 115
F- 14.0.8 VALVE #26 OKAY
a)
Remove the clamp from Valve #26!
b) Obtain a bucket!
c)
Per the Figure below
, remove the tubing from Valve #24’s INPUT and direct into the bucket
Figure 25 – Valve #24
d) From
Calibrate Hydraulics → Flow Pressure.
e) Press ‘CONFIRM’ THEN ENSURE the Flow Motor is rotating!
f) Measure from Valve #24’s tubing for one (1) minute!
g) Press ‘CONFIRM’ to return to “Calibrate Hydraulics”. More than 800 ml collected?
Yes
More than 800 ml! See procedure number
(page 115).
No
A)
Reconnect Valve #24’s tubing.
B)
TWO (2) possible bad components:
1)
Restricted
DiaSafe
©
filter OR
2)
Restricted tubing inside the DiaSafe
©
housing.
F- 14.0.81 GOOD FLOW FROM VALVE #24’S / ISOLATE ACTUATOR-TEST BOARD / VALVE #24
a) Reconnect Valve #24’s tubing.
b) Per the Figure above, touch valve 24’s (black) solenoid. Is it warm (i.e. energized)?
Yes
Solenoid is warm! TWO (2) possible bad components:
1)
Actuator-Test Board;
2)
Valve
#24.
No Solenoid in NOT warm!