2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 638
P- 2.0.0 INITIALIZATION DOES NOT COMPLETE / ISOLATE PUMPS
a)
IMPORTANT! To prevent damage, turn the machine OFF!
b) Figure below
unplug the ACTUATOR CABLE
from the distribution board.
c) Turn the machine on (fan running). Does System Initialization
reach 100%?
Yes
System Initialization reaches 100%! See
ISOLATING THE PUMPS
this page.
No
System Initialization does NOT reach 100%! See procedure number
(page 639)
ISOLATING THE PUMPS
Possible 24 volt ‘short circuit’ at the Acid, Bicarb, Heparin OR UF Pump.
A)
IMPORTANT!
Turn the machine OFF and reconnect the ACTUATOR cable.
B) Figures below, from the distribution board, unplug a pump (your choice) then continue to part C to see if
this is the problem.
C) Turn the machine on (fan running). Does System Initialization reaches 100%?
Yes
System Initialization reaches 100%! The unplugged pump may be causing the problem! To
confirm this turn the machine off, plug the pump back in, and turn the machine on. If System
Initialization does NOT reach 100% the pump is bad.
No System Initialization does NOT reach 100%! Turn the machine off and repeat parts B and C until
System Initialization DOES reach 100%. If after ALL pumps have all been unplugged and System
Initialization still does NOT reach 100% either the Actuator Cable OR the distribution board may
be bad.
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...