2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 36
F- 2.9.2 ISOLATE TEMPERATURE AND CONDUCTIVITY
a)
ENSURE NO ACTIVE LEAKS BEFORE CONTINUING!
b) Call the Home screen. Is [
Temperature
] REMAINING between 35.5 and 38.5 °C AND
Conductivity
REMAINING between 13.0 and 14.5 mS? TWO (2) possible scenarios:
1)
IF (and ONLY if) Temperature AND / OR Conductivity is
NOT
remaining in range:
Proceed
to
page 39
, procedure number F- 2.0.15.
2) IF Temperature AND Conductivity are in range:
See parts a THROUGH g below:
a) At the bottom of the screen, select the ‘Dialysate’ tab.
b) Figure right, if necessary, adjust the Conductivity
Limits until the ‘Actual’ Conductivity is CENTERED
between them.
c) Press ‘CONFIRM’!
d) Allow three (3) minutes OR until the Conductivity
window remains white!
e) Call debug screen 0 (Figure right) and locate
Valve #24’s ‘dot’. At this point it should be white.
(page 36).
F- 2.9.4 ISOLATE ‘OUT OF BYPASS’ CIRCUIT
a)
CLOSE THE SHUNT DOOR
!
b) Is Valve #24’s ‘dot’ BLUE?
Yes BLUE! See procedure number
(page 37).
No WHITE! See parts a THROUGH d below:
a) ENSURING the shunt door is FULLY closed, call debug screen 2.
b) ENSURE
CVRCLS
(2
nd
column from left) = 1!
c) Return to debug screen 0.
d) Allow up to three (3) minutes OR until if Valve #24’s ‘dot’ turns blue. TWO (2) possible
scenarios:
1)
IF (and ONLY if) it turns BLUE:
(page 37).
2) IF after three (3) minutes it stays WHITE:
Proceed to
page 38
, procedure number
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...