2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 317
SECTION 5 - CONDUCTIVITY / ON LINE CLEARANCE PROBLEMS
External leaks, “No Water” and Flow Errors cause Conductivity problems. If either occur at ANY TIME
address them first! To avoid error observe ALL stated times below!
A) ENSURE Dialysate Flow is ON (Flow on/off lamp NOT blinking)!
B) From the Home screen, ENSURE [Dialysate Flow] and has been
on and set to 800 ml/min for five (5) minutes!
C) If [
Conductivity
] is less than 12.6 mS, Figure right, ENSURE BOTH
connector plug O-rings are present! If not this may be the problem!
D) If the Automated Tests are running (screen reads “Test:.…”) allow them to finish.
E)
If installed, remove the ‘dummy chamber’ from the Level Detector
.
F) Select the [
Temperature
] window. Is “Temp Setting”
ALREADY = 37.0° C?
Yes
See part G.
No
Adjust “Temp Setting” to 37.0° C THEN continue to part G.
G) Press ‘CONFIRM’ to return to [
Temperature
].
H)
DO NOT RESET ALARMS!
I) Call debug screen 0. If
Flow Error
EVER
= 1, even just
once, there is a Flow Error! WITHOUT LOOKING AWAY
watch it for, TWO (2) possible scenarios below:
Scenario #1:
IF “Temp Setting” was already 37.0° C (it was NOT necessary to adjust it),
TWO (2)
MINUTES
OR until
if
Flow Error
EVER = 1.
Scenario #2:
IF “Temp Setting” was not already 37.0° C (it was necessary to adjust it),
FOUR (4)
MINUTES
OR until if
Flow Error
EVER = 1.
J) Is
Flow Error
is ALWAYS = 0?
Yes
Flow Error
ALWAYS = 0! See part K.
No
Flow Error
=
1 even if just once! Proceed to
page 20
K) Using a flashlight, ENSURE large air bubbles are NOT moving into the machine through the acid and
bicarb inlet tubing! If air is seen check the connector(s).
Continued on next page
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...