2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 186
Yes
BOTH ‘dots’ turn white for more than thirty (30) seconds again! If ABSOLUTELY SURE the
Dialysate Sampling
Hardware Option has a blue ‘X’ in the No box the Functional Board
may be bad.
No
BOTH dots DO NOT turn white at the same time. See parts a and b below:
a)
If the flow error was originally occurring in Heat Disinfect return to it! If not leave the
machine in RINSE.
b)
(page 186).
CLEAN- 5.3.0 VERR = 0 / ISOLATE TOTAL FLOW
a) Leaving the shunt door OPEN, ENSURE a “No Water” alarm NEVER occurs!
b) Watching for thirty (30) FULL seconds, does the external flow indicator’s ‘bob’ EVER rise about ½ way
up OR higher in the sight glass?
Yes
‘Bob’ moving! See procedure number
No ‘Bob’ NOT moving! See parts a through d below:
a) ENSURE no tubing restrictions at the Flow Pump AND the DiaSafe
®
filter.
b) ENSURE Flow Pump Pressure (gauge at Flow Pump) is PEAKING to between 34 and 37 psi.
c) ENSURE Loading Pressure (gauge in Rinse port) is peaking to between 22 and 27 psi.
d) Proceed to
page 562
to troubleshoot this restriction using procedure number
CLEAN- 5.4.0 ISOLATE FLOW ERROR
WITHOUT LOOKING AWAY, watch specifically for a “Flow Error” AND / OR a “Flow Inlet Error” for four (4)
minutes
OR
until if either one EVER occurs?
Yes
“Flow Inlet Error” AND / OR “Flow Error” occurs! Proceed to
page 187
, procedure number
No Neither alarm occurs! See parts a THROUGH c below:
a)
Close the shunt door!
b) Assuming neither alarm EVER occurs, allow
Remaining Prerinse Time
to reach 0:00!
c) Watch until if a
“Flow Recirc Error 1”
occurs OR until if the
program completes! TWO (2)
possible scenarios next page:
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...