2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 187
1)
IF (and ONLY if) “Flow Recirc Error 1” DOES NOT occur:
Whatever was causing
the Flow Error is no longer occurring! Do NOT continue to troubleshoot a Flow Error.
2)
IF “Flow Recirc Error 1” occurs:
(
page 175
CLEAN- 6.0.0 ‘BOB’ MOVING
/ ISOLATE VALVE #29 STICKING OPEN
a)
Turn the machine OFF!
b) Turn the machine on but
DO NOT
press CONFIRM!
c)
The “Select Program” banner MUST remain
up for this procedure!
d) Figure right, screw a 60 ml syringe, COMPLETELY
FILLED WITH AIR, onto the Fluid Sample Port.
e)
Per the Figure below
, locate Valves #29 and #30.
f)
Tightly clamp
Valve #30’s INPUT tubing.
g) Push hard on the syringe plunger while watching
Valve #29’s
OUTPUT tubing. Resistance is normal!
If Valve #29 is sticking open you will see air moving ALL THE WAY through it?