2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 149
NW- 2.0.1 VALVE ERROR = 0 / ISOLATE INLET WATER PRESSURE
a) Leaving the machine in a program, TURN THE WATER OFF!
b) A psi pressure gauge is required. ENSURE it reads 0 psi before installing it!
c) Per the Figure below, tee the gauge into Inlet Pressure Regulator #61.
d)
TURN THE WATER ON!
e)
WARNING! Using a plastic clamp here will cause error!
Using METAL PLIERS, tightly
clamp the tubing between the gauge and the Heat Exchanger at the location shown the Figure above.
TWO (2) possible scenarios:
1)
IF the gauge = 18 psi or more:
Remove the pliers then proceed to
page 151,
procedure
2) IF the gauge = less than 18 psi:
ENSURING the water was on prior to checking pressure, remove
the pliers then see procedure number
(page 149).
Figure 30 – Hydraulics Top View
NW- 2.0.2 ISOLATE INCOMING WATER
a) TURN THE WATER OFF!
b) This procedure requires a 1000 ml (or larger) graduated cylinder.
c) Reproduce the conditions when the “No Water” occurred. For
example, multiple machines running attached to the same
PORTABLE RO, filling the bicarbonate tank, etc.
d) Figure right, at the rear of the machine, remove the (TOP)
Incoming Water tubing!
Part e next page
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...