2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 438
FIL- 6.2.0 “OPERATOR ERROR” OCCURRED / TROUBLESHOOT DIALYSATE PRESSURE
a) ENSURE the transducer protector, at the external meter, is not WET or consider replacing it!
b) Return the Dialyzer lines to the shunt and close the door.
c) Turn the machine OFF.
d) Turn the machine on and return to Dialysis Program (“Select Program”
→‘Dialysis’ → ‘CONFIRM’)!
e)
IMPORTANT!
From the Home screen, select the [Dialysate Flow] window.
f) Set [Dialysate Flow] to 800 ml/min and press ‘CONFIRM’.
g)
Read before performing!
Return to (ABOVE) procedure number
page 437
) but, if
(and ONLY if) “Operator Error” reoccurs, one at a time, swap in the listed components (see
Component List below), and in between, return to procedure number
(page 437) until
“Operator Error” does not reoccur indicating the last component swapped in is the problem!
Component List
1)
Dialysate Pressure Transducer #9
a
;
2)
Sensor Board
b
;
3)
Actuator-Test Board;
4)
Functional Board;
5)
Sensor Board cable;
6)
Distribution board.
a
To LOCATE Transducer #9
refer to
(page 19).
b
To prevent “Cond Offset Failure” place the machine into
T and C Mode
(refer to
OPERATING
MODES
(page
Error! Bookmark not defined.
)
LEFT BLANK INTENTIONALLY
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...