
2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 475
PHT- 1.0.0 “GET NEG TMP” LASTING FOREVER / ISOLATE UF PUMP
A)
Call debug screen 1 to see
ATMP
(Figure right). It should
increase to about -250 during “Get Neg TMP” and if it does
“Neg Flow On” appears.
B)
If the “Get Neg TMP” banner is still up see part C. If NOT
allow the Tests to finish THEN,
even if they pass
,
proceeed to
page 476
, procedure number
C)
Figure right, remove the Fluid Sample Connector from its
port and hold it UP as shown!
D)
ENSURING the “Get Neg TMP” banner REMAINS up,
watch for up to TWO (2) minutes or until you see a strong
pulse that squirts out more than six (6) feet?
Yes
Strong pulse(s)! See procedure number
No
No strong pulse! TWO (2) possible scenarios 1) or 2) below:
1)
IF (and ONLY if) “Get Neg TMP” is still up:
Proceed to
page 503
2) ALL OTHER banners:
Allow the Tests to finish then,
even if they pass
, proceed to
page 476
, procedure number
PHT- 1.0.2 STRONG PULSE(S) SEEN
ENSURING “Get Neg TMP” remains up, watch
FILACT
(middle
column)
for one (1) minute. TWO (2) possible scenarios:
1)
IF FILACT EVER = 1!
A hydraulic air leak is indicated!
(page
2) IF FILACT ALWAYS = 0:
Is the “Get Neg TMP” banner still up?
Yes
“Get Neg TMP” still up!
(page 475).
No “Get Neg TMP” IS NOT up! Allow the Tests to finish then,
even if they pass
, proceed to
page 476
, procedure number
PHT- 1.0.3 “GET NEG TMP” STILL UP
a) Turn the machine off and remove the dummy chamber from the Level Detector module.
b) Turn the machine on and r
eturn to Dialysis Program (“Select Program” → ‘Dialysis’ → ‘Enter’)!
c) Proceed to
page 409
Summary of Contents for 2008K2
Page 1: ...2008K K2 HEMODIALYSIS SYSTEM TROUBLESHOOTING GUIDE Part Number 507298 Rev H ...
Page 655: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 651 Figure 132 Inside the Power Supply ...
Page 668: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 664 Figure 134 24 V Power Harness ...
Page 673: ...2008K K2 Troubleshooting Guide P N 507298 Rev H Page 669 APPENDIX C BLOOD PUMP ERROR CODES ...