2008K/K2 Troubleshooting Guide
P/N 507298 Rev. H
Page 117
F- 15.2.0 DEAP LESS THAN OR = 75 / ISOLATE DEAERATION MOTOR
Per the Figure below, is the DEAERATION MOTOR shaft rotating COUNTERCLOCKWISE (CCW)?
Yes
Rotating CCW! See procedure number
(page 117).
No NOT rotating CCW!
NOTE
ONLY
the DEAERATION MOTOR
will be checked then proceed to
page 130
Figure 26 – Deaeration Motor Rotating CCW
F- 15.2.1 DEAERATION MOTOR ROTATING CCW
Using the handle end
*
of a screwdriver, push HARD on and release the DEAERATION MOTOR’S shaft
several times. Can you make it stop rotating and REMAIN stopped?
Yes
If SURE the motor stops! TWO (2) possible bad components:
1)
Bad Deaeration Motor
(possibly brushes) OR;
2)
Bad Deaeration Pump head.
No
Motor continues to rotate! See procedure number
(page 118).